Sampai saat ini, lini produk kami mencakup

Orwell Pulse Oximeter Genggam

ORW-PO-002D

Handheld Pulse Oximeter is a device that measures oxygen saturation, pulse rate and patient perfusion index that can be used inside and outside the hospital environment. This device is designed with high flexibility and portability to provide fast, accurate and reliable measurements.

Optional Probe
Spesifikasi Produk
  • 3.2-inch high-definition LCD.
  • Large screen to display SPO2/PR/PI/Pulse Bar Graph/Plethysmography.
  • Available in multiple languages ​​including Indonesian and English.
  • Equipped with external oximeter probe:
    • Adult
    • Child
    • Neonatal
  • Equipped with accessories:
    • 1 bag
    • 1 unit data cable (USB type C)
    • 1 unit extension cable
    • 1 unit adapter
  • Equipped with rechargeable Lithium battery reset and battery indicator.
  • The battery can be used for measurements for 8-11 hours.
  • Equipped with a USB port for charging and communication with external devices.
  • The measurement data is transferred to a laptop or computer for reading and analysis.
  • Measurement parameters:
    • SpO2, PR, PI, PVI, and RR
  • Measurement Range:
    • SpO2: 0% - 100%, Accuracy: 70% - 100% or ± 2% (resolution ± 1%)
    • PR: 25 bpm - 250 bpm, Accuracy: ± 2 bpm or ± 2%
    • PI: 0 - 20% (resolution 0.001%)
    • RR: 0 - 70 bpm, Accuracy: ± 2 bpm (resolution ± 1 rpm)
    • PVI: 0 - 100% (resolution ± 1%)
  • Audio and visual alarm function for indication:
    • SpO2 and pulse rate limits
    • Low battery power
    • Incorrect probe installation position on the patient
  • High accuracy and reliability at blood perfusion 0.05%
  • Very suitable for use inside and outside the hospital environment.
  • Can store and monitor patient data trends during 360 hours even data will not be lost when the power is off.
  • According to SNI ISO 80601-2-61:2014 and SNI IEC 60601-1:2014 standards